Medical Device Maintenance
31 December 2022 2024-12-15 18:28Medical Device Maintenance
Medical equipment maintenance
Plan the maintenance of your medical devices
To reduce the likelihood of medical devices malfunctioning, periodic preventive maintenance must be carried out.
Our preventive inspection services are carried out according to predetermined technical criteria, specified in the manufacturers’ operating and service manuals.
Each inspection includes a simulator test, a general function test and, for some devices, an electrical test. This also includes checking consumables to ensure correct use of the medical device on patients.
Preventive maintenance also aims to reduce the risk of breakdowns and maintain equipment performance over time, to ensure safe operation.
Swixed produces a report for each device inspected. It records the values given by the device during simulation or electrical testing. It may or may not certify that the system is working properly.
Submit the form below to request a free quote for all your medical devices.
Maintenance services
Inspection and maintenance
Maintenance work is carried out by our team of technicians, on whom you can rely at all times. Below is a list of tests performed on each type of device.
Maintenance et contrôle des Défibrillateurs AED
- Testing general device functions
- Arrhythmia detection simulator test (coarse ventricular fibrillation (VFBC), fine ventricular fibrillation (VFBF), ventricular tachycardia (VTAC), atrial fibrillation (AFIB))
- ECG simulator test at different heart rates
- Simulator test of electric shock and measurement of the number of joules delivered
- Electrode control and functionality
- Check the battery and measure its charge level if possible.
- Data extraction where possible.
- Report writing. Indicating whether the values emitted by the device are within tolerances, and whether accessories and consumables are still functional and can remain so until the next inspection.
Maintenance et contrôle des Électrocardiographes
- Testing general device functions
- Electrical grounding and leakage current test according to IEC 60601-1.
- Simulator test of the electrocardiogram printout. Measurement and control of printed layouts.
- Check that the patient cable is working properly.
- Battery check.
- Check mains cable or charger and fuses.
- Cleaning and maintenance of the print roller.
- Report writing. Indicating whether the values emitted by the device are within tolerances and whether the accessories are still functional. Printout of successful electrical test report.
Maintenance et contrôle des Moniteurs patients
- Testing general device functions
- Electrical grounding and leakage current test according to IEC 60601-1.
- ECG – Simulator test of electrocardiogram printout. Measurement and control of printed layouts. Check that the patient cable is working properly.
- NIBP – Simulator-based calibration of systolic, diastolic and pulse values. Has several different values. Calibration if device permits. Leak test of cuff(s) supplied.
- SPO2 – Calibration of pulse oximeter with oxygen saturation simulator and pulse. Has several different values. Check spo2 sensor(s).
- Battery check.
- Check power cable or charger and fuses.
- Maintenance cleaning of the print roller on the integrated printer (if available).
- Report writing. Indicating whether the values emitted by the device are within tolerances and whether the accessories are still functional. Printout of successful electrical test report.
Maintenance et contrôle des Remler/Mapa 24h
- Testing general device functions
- Simulator calibration of systolic, diastolic and pulse values. Has several different values.
- Calibration of tensiometer (manometer) using a reference manometer
- Supplied cuff leak test.
- Internal battery check
- Checking rechargeable batteries
- Data transfer test from device to PC
- Boso ProfilManager software update if required
- Report writing. Indicates whether the values emitted by the device are within tolerances and whether the device can still be used.
Service d’étalonnage des tensiomètres électroniques
- Complete check of functions and stability.
- Calibration of systolic, diastolic and pulse pressure values to different reference values to ensure optimum accuracy.
- Cuff tightness check to ensure reliable measurements under normal conditions of use.
- Delivery of a detailed report specifying compliance with accepted tolerances, and recommendations for continued use.
Service d’étalonnage pour tensiomètres manuels
- Pressure gauge calibration: Use of a simulator and a reference pressure gauge to guarantee accuracy.
- Leak test: Checking the cuff to ensure it is working properly.
- Report writing: A detailed report indicating the conformity of the measured values and the general condition of the device will be supplied.
Étalonnage des oxymètres de pouls
- Calibration of pulse oximeter with oxygen saturation simulator and pulse. Has several different values.
- Report writing. Indicates whether the values emitted by the device are within tolerances and whether the device can still be used.
Form
Maintenance request
Please complete this maintenance request, detailing the equipment (make and model) for which you would like to carry out maintenance. You will receive a free quote for your medical devices as soon as possible.