Edan F3 twin fetal monitor – 2-channel cardiotocic RCF
CHF 2'999.00CHF 3'468.00
The Edan F3 Twin Fetal Monitor, or cardiotocograph, is a lightweight device (3.5 kg) with 2 RCF channels and 1 TOCO channel. Equipped with a color screen and IPX8 waterproof sensors, it ensures precise fetal monitoring thanks to its high sensitivity (1 MHz). For home use, it offers advanced features for optimal monitoring of twin pregnancies.
Description
Brand
Edan
Additional information
Weight | 3.5 kg |
---|---|
Dimensions | 35 × 30 × 10.4 cm |
Screen | 5.6-inch TFT LCD color // Resolution 640 (width) x 480 (height) // Plot and number display |
Type | Analysis of fetal heart rate and monitoring of intrauterine growth retardation (IUGR), with calculation of STV (Short-Term Variability), including SOV technology for twin pregnancies. |
Keyboard | Silicone keypad for easy monitoring |
Indicators | Signal quality indicator helps optimize sensor position |
Ultrasound - Technology | Ultrasonic pulsed Doppler with auto-correlation |
Ultrasonic - Repetition rate | 2 KHz |
Ultrasonic - Pulsation | Duration: 92 s |
Ultrasonic - Frequency | 1 MHz ±10 |
Ultrasonic - Measuring range | 50 bpm-240 bpm |
Ultrasonic - Resolution | 1 bpm |
Ultrasonic - Precision | ± 2 bpm |
Toco - Measuring range | 0% – 100% |
Toco - Resolution | 1% |
Toco - Zeroing | Automatic / Manual |
Toco - Non-linear error | 10% |
Fetal ECG - Technology | Peak-to-peak detection |
Fetal ECG - Measuring range | 30 bpm-240 bpm |
Fetal ECG - Resolution | 1 bpm |
Fetal ECG - Accuracy | ± 1 bpm- |
Fetal ECG - CMRR | >110 dB-. |
Fetal ECG - Noise | < μVP |
Fetal ECG - Impedance | >10 M (differential, DC 50/60 Hz) // >20 M (common mode) |
PlU - Intrauterine pressure - Measuring range | 0mmHg – 100mmHg |
PlU - Intrauterine pressure - Resolution | 1% |
PlU - Intrauterine pressure - Zeroing | Automatic / Manual |
PlU - Intrauterine pressure - Non-linear error | ±3 mmHg |
Automatic Fetal Movement - Technology | Ultrasound Doppler |
Automatic Fetal Movement - Measuring range | 0% – 100 |
Automatic Fetal Movement - Resolution | 1% |
Manual Fetal Movement - Method | Remote marker |
Registration - Type | Thermal printer |
Registration - Paper | Thermo-sensitive paper (150mm) |
Recording - Speed | 1/2/3 cm/min (real time) |
Registration - Resolution | 8 dots/mm |
Registration - Information | RCF1 trace, RCF2 trace, TOCO trace, MFA trace, fetal movement mark, event mark, aut-zero symbol, date, time, recording speed, ID, name, RCF2 offset, etc. |
Power supply | 100 V- 240 V |
Refill | 14.8 V rechargeable lithium battery |
Power | Battery capacity: 4200 mAh/2100 mAh |